MGas NH3 5.1

 

 

 

FUNCTION

The instrument for measuring the following Gas- Concentrations
• Ammonia %NH3

Customizable Gas components are possible.

Additional measurement:
– Temperature Based on state-of-the-art Tunable Diode Laser Spectrometry (TDLS) with high accuracy, good long-time stability, and excellent repeatability, this equipment is qualified to measure furnace atmospheres online.

Eight analogous outputs, eight digital inputs/outputs, and an optional serial interface are possible. All Analog and Digital inputs/outputs can be configured manually e.g. as Alarm, starting calibration, or switching off-pump. An optional data logging function with USB data transfer and internal memory with 16MB is available. The logging function can work in manual or timer mode. With the delivered Software “MGas Viewer” you can visual, print, export, and administrate your stored data. The equipment is adjustable by a high- quality foil keyboard with self explainable menu structure in German and English language. Automatic calibration for zero-point and span for all gas components is possible. Additional it is possible to configure several protection functions, e.g. low process temperature, in order to increase the lifetime of the device.

Gas Analyzer System MGAS 5.x

TECHNICAL DATA:

 

Dimensions:
  • Desktop variant: 450 x 200 x 440 (w x h x d)
  • Rack variant: U=4, HP=84, D=400mm
Weight: Approx. 15 kg
Protection type: IP 20 to IEC 529
Power supply: 230V +4% / -10%, 50-60Hz or
115V ± 10%, 50-60Hz
Power: Approx. 100W / Slow Fuse 2A in Power switch
Heat up time: Approx. 15 min.
Climate: Storage: 0…50 °C
Operation: 5…40 °C
Measuring range: NH3: 0…20.000,0 ppm
Measuring gas components: NH3
Measuring method: TDLS – Tunable Diode Laser Spectrometry

 

 

Additional measurement: Temperatur Taupunkt
Thermocouple K, S
Influence values: Accuracy: ± 2% of reading or, ± 2 ppm whichever is larger
Precision 2σ : ± 0.9 ppm
Linearity: included in the accuracy
Repeatability: included in the accuracy
Zero drift over 2 h period:: below the accuracy
Span drift over 8 h period: below the accuracy
Operating pressure: 800 … 1100 mbar
Gas inputs:
  • Measurement gas input
  • Zero gas input
  • Reference gas input
Pressure:
  • Maximal pressure: 100 mbar overpressure (ü)
  • Working pressure : 0 do 50 mbar overpressure (ü)
  • Gas flow: Approximately 3 l/min
Gasausgänge: Gas output 1
Gas output 2
Pressure: Pressureless